Faulty Phenomenon Malfunction: the device responds to the power button but fails to start the ultrasound application. A malfunction was detected in the hard disk. The hard disk was replaced and the... ver más
Faulty Phenomenon Problem description: when turning on systems, an error 005 (005. 3DR4. 692A. 16) appears. Troubleshooting and Solutions Description of the problem: when turning on the system, an... ver más
Faulty Phenomenon A stripe appeared on the display of the RST board (PN: KTZ196358, KTZ280122). RST - impulse transmission board, 16 pieces are installed in the device (1 faulty). Sent the board... ver más
Faulty Phenomenon GE Logiq E10: A malfunction was detected in the TX board, an element that controls the switching on/off of high voltage for supply to the acoustic module has failed. The device is... ver más
Faulty Phenomenon Beamformer consists of RFM201, RFM221 (PN: KTZ303915, KTZ303916); RSX10, RSX20 (KTZ303250, KTZ303054). Error: Aurora communication failed; when turning on the device, the image... ver más
Faulty Phenomenon Repair block of power supply Main Power Supply (PN: 5205054). Malfunction: absence of 48V for the computer power supply block and therefore the device does not turn on. ... ver más
Faulty Phenomenon Malfunction: Memory access violation, device requires restart. Repair board RFM423G. P19 (PN: KTZ304074, KTZ304142): restoration of factory firmware in EEPROM and Flash Memory for... ver más
Faulty Phenomenon Repair the Main Power Supply (PN: GA200730-5) power supply block. Malfunction: when turning on the device, it shows the error "System Error". The logs indicated that the ultrasound... ver más
Faulty Phenomenon Repair the Main Power Supply (PN: 5205052-7) power supply block. This power supply unit is also suitable for Vivid E9. Malfunction: does not turn on the device, no 48 V on the... ver más
Faulty Phenomenon Repair RFM423 (PN: KTZ304074) of the GE Voluson E10 (versions BT17, BT18, BT19, BT20, BT21) device. Malfunction: when turning on the probe, the image begins to display intense... ver más
Faulty Phenomenon Repair Acquisition Module for A. X hardware (PN: 453561704245) of the Philips Epiq 5/7 device. Malfunction: During the examination in all modes of operation and on all sensors,... ver más
Faulty Phenomenon Repair flat panel DBM128G ASSY (PN:5573638-2) from GE Voluson S10 BT18. Malfunction: when turning on the device, the image begins to intensively display white noise across the... ver más
Faulty Phenomenon Repair the motherboard of the Philips Epiq 5G device malfunction: the device does not turn on and does not respond to the power button. Troubleshooting and Solutions Repair of... ver más
Faulty Phenomenon GE VIVID I Ultrasound System Repair Case: Fixing "Probe Not Recognized" Error Troubleshooting and Solutions GE VIVID I Ultrasound System Repair Case: Fixing "Probe Not... ver más
Faulty Phenomenon Repair power supply module (APS Assy PN: 5166108) from GE Logiq P5/A5 device malfunction: device fails to start, no ultrasound imaging, no high voltage, no indicator light. ... ver más