SIEMENS Polydoros LX 30/50 X-Ray Generator

Descripción

Especificaciones

Model Rating

Not Rated
Rate This Model Now
Information Provided By:
Product-specific Remarks 0
Safety information 0
NOTE When carrying out the work steps and checks, the product-specific
safety information contained in the documents. as well as the
general safety information must be observed.
Protective measures 0
• Prior to performing any activity in the generator, switch it off at the power OFF switch on
the D160 board.
WARNING With the generator switched off, there is still line voltage at the T1
transformer and on the D160 switch-on circuit. After switching off
the generator, there is still approx. 600 V DC voltage present for
the inverter!
􀂹 This is indicated by LEDs V35 and V36 on the D110 and
LED V89 on D220 lights up. The voltage dissipates within
approx. 1.5 minutes to 0 V; the LEDs go off at approx. 30
V.
• To switch power off to all parts of the system (generator and connected components),
set the system switch to the OFF position.
• To prevent unintentional triggering of high voltage or radiation, set the SS switch (S1)
on the D100 to OFF (no control of the inverter).
• Install or remove assemblies
Partial List - Error Code's
Error 001 0
Designation:
Start of a service session.
Description:
• The Service PC has successfully contacted the XCU.
• Not an error; for information only.
• This error appears in the error log under the Number 41 from 240 (XCU).

Error 402 0
Meaning:
Minimum filament current
Description:
The minimum filament current (=1/2 nominal value) is monitored in standby.
Possible causes and action:
• Intermediate circuit voltage missing in the filament circuit at X41.7 and X41.10, mearure
230V ± 10%.
• Check fuse F21, D160
• Interruption in the filament path:
- Check wiring X41 - H1
- Workstation selected, AP relay in H1 actuated
1. H1 test point 50 = 0V
2. Ap 1 test point 51 = 24V (±15%)
3. AP2 test point 52 = 24V (±15%)
• Check filament transformer primäry winding (R < 1Ohm, transformation ratio 32:9)
• Check filament and high-voltage cable
• If no error is found, replace board D220.

Error 610 0
Meaning:
Invalid range of tube current (10 mA ADJUSTMENTS --> GENERATOR PARAMETER).
If this is done with generator, using older D100 SW, this error might be caused,
especially in Tomography. Set current back to 10 mA.

Error 421 0
Meaning:
Wrong tube current value in fluoroscopy
Possible causes and action:
Error in the master → perform system configuration and POLYDOROS adjustment.





 

Reviews

Forums

Share your opinion or experience with this equipment - Review this model

RaySafe helps you avoid unnecessary radiation

RaySafe solutions are designed to minimize the need for user interaction, bringing unprecedented simplicity & usability to the X-ray room. We're committed to establishing a radiation safety culture wherever technicians & medical staff encounter radiation.

Anuncie
Aumente su conciencia de marca
Subastas + ventas Privadas
Consigue el mejor precio
Comprar Equipo/Piezas
Encuentra El Precio Más Bajo
Noticias diarias
Lee las últimas noticias
Directorio
Examina todos los usuarios DOTmed
Ética en DOTmed
Ver nuestro programa de ética
El oro parte programa del vendedor
Recibir las solicitudes de PH
Programa de distribuidor con servicio gold
Recibe solicitudes
Proveedores de atención de salud
Ver todos los HCP (abreviatura de asistencia médica) Herramientas
Trabajos/Entrenamiento
Encontrar/rellenar un trabajo
Parts Hunter +EasyPay
Obtener presupuestos para piezas
Certificado recientemente
Ver usuarios certificados recientemente
Recientemente clasificado
Ver usuarios certificados recientemente
Central de alquiler
Alquila equipos por menos
Vende equipos/piezas
Obtén más dinero
Mantenga el foro de los técnicos
Buscar ayuda y asesoramiento
Petición sencilla de propuestas
Obtén presupuestos para equipos
Feria comercial virtual
Encuentra servicio para el equipo
El acceso y el uso de este sitio está conforme a los términos y a las condiciones de nuestro AVISO LEGAL & AVISO DE LA AISLAMIENTO
Característica de y propietario DOTmeda .com, inc. Copyright ©2001-2019 DOTmed.com, Inc.
TODOS LOS DERECHOS RESERVADOS