centrifuge 5804r showing error 3
August 21, 2014 12:51

Error 3 Cause no rotor No rotor pulses during run or acceleration. Missing rotor, measuring system for rotational speed defective, drive defective. – Rotor not fastened. – No pulses after start. Device brakes for 15 ...Click here to read more

centrifuge 5804r showing error 3
August 21, 2014 12:51

Here is what the service manual says about error 3 Error 3 no rotor No rotor pulses during run or acceleration. Missing rotor, measuring system for rotational speed defective, drive defective. – Rotor not fastened. – No...Click here to read more

Heraeus MEGAFUGE1.0
December 10, 2012 06:34

What do you mean by RPM problem? What, exactly, is the unit doing? ...Click here to read more