The T9/T10DCS Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy.
The optical design of both the T9DCS and T10DCS offer extremely low stray light characteristics (≤0.00004%T NaI,220 nm) which allows for an extensive photometric range (-8.0 – 8.0Abs). Measurements at deep ultra-violet wavelengths can also be achieved with use of Nitrogen purged optics.
The instrument can be optically configured to suit the needs of the sample by means of a continually adjustable slit for precise control of spectral resolution and beam size adjustment by means of an attenuating wheel.
Key Features
◆ True Double Beam Double monochromator Optics, and fully sealed to provide ultra-low stray light (0.00004%%T @220nm) and high photometric range (-8.0~8.0 Abs)
◆ 185– 900nm Wavelength range with Nitrogen purge
◆ Precise Wavelength accuracy is ensured by the integrated Mercury lamp for automatic correction
◆ Continuously selectable spectra bandwidth from 0.1-5nm
◆ Beam size continuously adjustable for variety of applications including ultra-volume cuvettes
◆ Large sample compartment suitable for all kinds of accessories, especially for 150mm integrating sphere
◆ Support WIFI communication in field operation
◆ The integrated Mercury Emissions Lamp with automatic correction function ensures precise wavelength accuracy
Specifications
Model T9DCS T10DCS
Optical System
→Click here to know more about
Optical Path System←
Double beam Double beam
Scan Speed 1000 nm/min
1500 nm/min location speed 1000 nm/min
1500 nm/min location speed
Wavelength Range 185-900 nm 185-900 nm
Wavelength Accuracy ±0.2 nm ±0.2 nm
Spectral Bandwidth
→Click here to know more about
SBW←
±20 % (0.1 nm-5 nm continuously adjustable)
±20 % (0.1 nm-5 nm continuously adjustable)
Photometric Mode
Transmittance,Absorbance,
Enegy and Concentration
Transmittance,Absorbance,
Enegy and Concentration
Photometric Range -6.0 Abs-6.0 Abs -8.0 Abs-8.0 Abs
Photometric Accuracy ±0.004 A@2.0 A
±0.003 A@1.0 A
±0.002 A@0.5 A
±0.3 % ±0.004 A@2.0 A
±0.003 A@1.0 A
±0.002 A@0.5 A
±0.3 %
Photometric Reproducibility ≤0.002 A@2.0 A
≤0.0008 A/1.0 A
≤0.0004 A/0.5 A
≤0.1 % ≤0.002 A@2.0 A
≤0.0008 A/1.0 A
≤0.0004 A/0.5 A
≤0.1 %
Photometric Noise 0 % Noise: ≤0.01 %
100 % Noise: ≤0.1 % 0 % Noise: ≤0.01 %
100 % Noise: ≤0.1 %
Baseline Flatness ±0.0008 Abs ±0.0005 Abs
Baseline Stability 0.1 %/hr 0.1 %/hr
Stray light
→Click here to know more about
Stray Light←
≤0.0001 %T (NaI, 220 nm)
≤0.0001 %T (NaNO2, 360 nm) ≤0.00004 %T (NaI, 220 nm)
≤0.00002 %T (NaNO2, 360 nm)
Cell Holder Automatic 8 Cell Changer Automatic 8 Cell Changer
Detector Photo multiplier tube Photo multiplier tube
Light Source D2 lamp,Tungstern lamp,Mercury lamp for WL correction D2 lamp,Tungstern lamp,Mercury lamp for WL correction
PC Interface RS232, USB, Wifi RS232, USB, Wifi
Software Support UV Win UV Win
Power Supply Switchable 120-230 VAC 50-60 Hz Switchable 120-230 VAC 50-60 Hz
Weight 136 lb 136 lb
Dimensions
(Width, Depth, Height) 36 x 25 x 12 in 36 x 25 x 12 in