We have a Horiba ABX Pentra 500 Chemistry Analyzer For Sale
ABX Horiba Pentra 400 for sale
Sold As is, has been tested, in great working condition.
Includes: ( All itmes shown in photos)
- Analyzer Unit
- Keyboard and mouse
- Cables
- Battery unit
Also available seperately:
- Refurbished
- Re-certified
- Service Agreement
- Preventive maintenance.
- Installation
- Extended Warranty
And more! Please Contact us for more information.
-.
From routine to specific tests, perform them all on a single system.
Up to 420 tests per hour with ISE module, the ABX Pentra 400 is an extremely compact Clinical Chemistry benchtop analyser.
Its great autonomy with continuous loading provides enchanced productivity in a user-friendly environment.
Overview
55 on-board assays
Innovative cassette format
Integrated work station and validation station
Touch screen
Real-time viewing of your workflow
Features
General Description
Analysis Method Types:
Spectrophotometry: Colorimetry and Turbidimetry
Potentiometry: Direct (Serum or Plasma) and Indirect (Urine)
Up to 55 assays on board
Optional ISE Module (Na+, K+, Cl-)
Maximum Throughput: 420 test/h
FDA cleared
Sample Handling
Sample Capacity: 6 racks with 10 sample positions for primary tubes or cups
Continuous loading
Positive Sample Identification: Integrated Barcode Reader and Tube Presence Detector
Sample Integrity:Level, Shock and Clot Detection
Sample Collection Format: 0.7, 4, 5, 7 and 10 mL (diameter: 16mm)
Sample Specimen: Serum, Plasma, Urine, CSF, Whole Blood, Homogeneous Fluids
Sample Volume: 2 to 380 µL
Automatic Sample Dilution: from 1/2 to 1/22,500
ISE: 10 µL (Indirect) or 65 µL (Direct)
Automatic Post Concentration: from x2.0 to x 10.0
Automatic Re-run
Calibrator/Control Handling
Calibrator/Controls Positioning on Sample or Reagent Tray
Automatic Standard Dilution Series:
Dilution Factors: Main Direct 1/2 - 1/150
Dilution Factors: Main Indirect 1/2 - 1/1508
Reagent Handling
52 Reagent positions on board:
44 in a closed refrigerated area (2-10°C)
8 at room temperature
Positive Reagent Identification:Integrated Barcode Reader
Reagent Probe: Level Detection and Preheating
Reagent Management with Cassettes: Automated Reagent On-board Stability Check plus Test Countdown with automatic back-up
Reagent Volume: from 2 to 600 µL
Reaction System
Reaction Volume: 150 to 600 µL
Reaction Temperature: Air bath at 37°C
Photometer Reaction Cycle:12 seconds
Mixing with Teflon-coated Stirrer
Reaction Time: 1 to 100 cycles
Disposable Acrylic Cuvettes
Automatic Cuvette Changer
Cuvette Capacity: 36 segments (432 cuvettes)
7 Integrated Calculation Algorithms
This item is sold as is with everything included as shown in photos
SPECIFICATIONS
Physical:
System Size: 40” x 28” x 25” (W x L x H)
System Weight: 264 lbs
External Cooling Unit Size: 12” x 17” x 16” (W x L x H)
External Cooling Unit Weight: 77 lbs
Printer Printer
Sound level: < 60 dBa
Operating temperature: from 15 to 32°C
Operating Humidity: from 20 to 85%
Total Power Requirements (with Printer) 1000 VA
Water De-ionized/Distilled Water - 10L Tank
Water specifications Resistivity > 5 MegOhms
Conductivity < 0.2 µS/cm
Computer & Sofware
Operating system: Windows XP
Data processing: 12 inch Color LCD Touch Screen
Database Capacity: unlimited results
Processor: 600 MHz
RC 232C: Patient results, QC to LIS with ASTM protocol
Integrated Validation Station
Quality Control Management Levey Jennings, Automatic QC, Default controls and Westgard rules
Logs Reagents, Quality Control, Calibrations
Applications, Maintenance, Sequencing, Flags and Alarms
Export of archived data and Quality Control data (QCP) USB key
Audible alarms
Interactive Help HTML Format
Optical System
Tungsten-halogen Lamp
Measurement Principle Bi-chromatic light absorbance
Choice of 15 wavelengths from 340 to 700 nm
Wavelength selection with holographic grating technology
Sensor Photo diode array
Optical Linearity 2.5 A. at 340mm; 3.0 at other wavelengths
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.