STAT FAX 1904 Classic Biochemistry Analyzer
Programmable open system
•Read well accepts 12 mm test tubes
•Standard 6 filter configuration (340 nm - 600 nm)
•Reads and Calculates the results of endpoint and kinetic colormetric clinical assays
•Kinetic Assays may be plotted on the internal graphics printer to verify linearity of the reaction
•Built-in 37˚C incubation block with 12 stations
•Single or multipoint calibration
•Stores more than 50 tests in memory
Specifications
Photometer
Stability Drift of no more than 0.005A in 8 hours/bichromatic
Light source Tungsten halogen lamp, with lamp saver feature
Standard wavelengths 340, 405, 450, 505, 545, 600 nm
Optional filter range 340 nm - 700 nm (custom order)
Reading range 0 to 2.5 absorbance units
Sampling
12 mm tube, minimum sample volume 1 ml
Read speed 3 seconds per tube
Temperature 37˚C, block with 12 stations is constantly on, read cell has an on/off switch and may be disabled as required
Operating Modes
Calculation modes Single point calibration by standard or factor, multipoint calibration with point-to-point curve fit, rate by standard or factor (batch or singly)
Test menu More than 50 open channels to store tests. Stores all parameters including wavelengths, calculations, unit codes, linear and normal ranges, rate timing, standard values, test names and previous standard curve
Dimensions 9” (22.86 cm) x 13.5” (34.29 cm) x 5” (12.7 cm)
Weight (10 lbs) 4.5 kg
Power 115/230 V AC 50/60 HZ (switch selectable)
Return Policy