Repaired:
GE HV Cable p/n: 00-901141-04 or 00-901141-01 / 5392467FS / 5428746FS / 5694681-01
GE HV Cable p/n: 00-901141-05 or 00-901141-03 / 5694681-02
GE HV Cable p/n 5694683 for Motorized unit.
HV TESTED AT 250kv FOR LEAKAGE, from pin to pin to check for heat cracks in wire jacket as the three wires are twisted together throughout each of the two High Voltage cables.
If there are any heat cracks in the Anode and Cathode wire insulation, our 250kv test will find it.
Anode and Cathode pin lengths are verified to eliminate the low Ma issues.
Entire cable tested on PC based tester to verify all connections after final cleaning of assembly.
In stock and ready to ship.
CORE RETURNS ARE MANDATORY
We will not accept what we consider "non-repairable" cables as core exchanges. In that case, ask for Outright price.
90 Day warranty
Always buying CLEAN HV Cable cores that can be repaired.
If you have any to sell, contact us for details.
Separation of Jacket at the mounting block may be repairable. Need good video cables, good Anode and Cathode.
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.